MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ENH PLT PLS RBC AUTORAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ENH PLT PLS RBC AUTORAC Back to Search Results
Model Number 80360
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly a result of an escape of wbc from the lrs chamber during the latter portion of platelet collection.Based on the available information, it cannot be ruled out that the higher-than- expected wbc content in the platelet product could be donor-related.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(4).The platelet collection set is not available for return because it was discarded by the customer.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: the signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly a result of an escape of wbc from the lrs chamber during the latter portion of platelet collection.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ENH PLT PLS RBC AUTORAC
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key10212177
Report Number1722028-2020-00316
Device Sequence Number3093644
Product Code GKT
UDI-Device Identifier05020583803602
UDI-Public05020583803602
Combination Product (Y/N)N
Initial Reporter CountryPR
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2021
Device Model Number80360
Device Catalogue Number80360
Device Lot Number1912202130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2020
Supplement Date Received by Manufacturer01/26/2021
Initial Report FDA Received Date06/30/2020
Supplement Report FDA Received Date02/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-