Model Number 5MAXJETDKIT
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional 510(k)# that also applies to this complaint: (b)(4).The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2020-00848.
 
Event or Problem Description
On (b)(6) 2020, the patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a neuron max 6f 088 long sheath (neuron max), a neuron select catheter 6f, a penumbra system jetd reperfusion catheter (jetd) and a velocity delivery microcatheter (velocity) and a non-penumbra sheath.During the procedure, the physician placed a sheath with the neuron max and 6f select.Then, the physician advanced the jetd over the velocity through the neuron max and experienced resistance at the distal end of the neuron max; therefore, the entire system was removed.After the removal of the entire system, it was noticed that the neuron max was kinked.The procedure was completed using a new penumbra system ace 60 reperfusion catheter (ace60), a new neuron max, and the same velocity.There was no report of an adverse effect to the patient.On (b)(6) 2020, a post-procedure imaging was performed, and it indicated that there was an unknown item remaining in the patient¿s brain.It was reported that the item had broken off and migrated into the open m2 segment of the mca due to the other branch of m2 being occluded.No intervention was performed.Patient¿s status unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM JETD REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10215429
Report Number3005168196-2020-00849
Device Sequence Number2565216
Product Code NRY
UDI-Device Identifier00814548019802
UDI-Public00814548019802
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K090752
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/02/2022
Device Model Number5MAXJETDKIT
Device Catalogue Number5MAXJETDKIT
Device Lot NumberF93456
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2020
Initial Report FDA Received Date06/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Date Report Sent to FDA01/01/2005
-
-