| Model Number |
PCB00 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/ date of birth: unknown, information not provided.Gender/ sex: unknown, information not provided.Date of event: unknown, not provided.If implanted, give date: unknown, information not provided.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that lens became displaced after omni procedure and surgeon decided to explant the lens from patient¿s operative eye in secondary surgical procedure.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: device available for evaluation; returned to manufacturer on: 7/28/2020.Device evaluation: the product was returned.The lens associated to the complaint was received in a specimen container; the pcb00 insertion device was not received.The lens is cut most probably related to the reported removal.Both haptics are complete, there is no damage on the haptics.The reported issue cannot be verified; there is no indication of a product quality deficiency.Based on the product returned evaluation a product quality deficiency or product malfunction could not be determined.Manufacturing records review: the manufacturing records for the device were reviewed.The device was manufactured and released according specification in compliance with the product intended use as required.A search revealed no additional investigation request (ir) for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|