• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/ date of birth: unknown, information not provided.Gender/ sex: unknown, information not provided.Date of event: unknown, not provided.If implanted, give date: unknown, information not provided.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that lens became displaced after omni procedure and surgeon decided to explant the lens from patient¿s operative eye in secondary surgical procedure.No further information provided.
 
Additional Manufacturer Narrative
Additional information: device available for evaluation; returned to manufacturer on: 7/28/2020.Device evaluation: the product was returned.The lens associated to the complaint was received in a specimen container; the pcb00 insertion device was not received.The lens is cut most probably related to the reported removal.Both haptics are complete, there is no damage on the haptics.The reported issue cannot be verified; there is no indication of a product quality deficiency.Based on the product returned evaluation a product quality deficiency or product malfunction could not be determined.Manufacturing records review: the manufacturing records for the device were reviewed.The device was manufactured and released according specification in compliance with the product intended use as required.A search revealed no additional investigation request (ir) for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10216010
Report Number2648035-2020-00461
Device Sequence Number2552393
Product Code HQL
UDI-Device Identifier05050474558212
UDI-Public(01)05050474558212(17)221105
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/05/2022
Device Model NumberPCB00
Device Catalogue NumberPCB0000175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/03/2020
Supplement Date Received by Manufacturer07/07/2020
10/25/2020
Initial Report FDA Received Date06/30/2020
Supplement Report FDA Received Date07/31/2020
10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-