SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Mechanical Jam (2983)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The sample is under evaluation by the manufacturing site.
 
Event Description
It was reported that the motor was not working.No case involved.Preliminary results of investigation have concluded that this unit had a blade stall which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issues.A functional evaluation revealed a hand piece stall.The complaint was confirmed and the root cause has been associated with a mechanical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.H11: information on b5 corrected.
 
Event Description
It was reported that the motor was not working.No case involved.Results of investigation have concluded that this unit had a blade stall which makes it a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10216169
MDR Text Key197094277
Report Number1643264-2020-00446
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public3596010555472
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2020
Initial Date Manufacturer Received 06/05/2020
Initial Date FDA Received06/30/2020
Supplement Dates Manufacturer Received07/07/2020
Supplement Dates FDA Received07/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-