Catalog Number 72200616 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The sample is under evaluation by the manufacturing site.
|
|
Event Description
|
It was reported that the motor was not working.No case involved.Preliminary results of investigation have concluded that this unit had a blade stall which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H3, h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issues.A functional evaluation revealed a hand piece stall.The complaint was confirmed and the root cause has been associated with a mechanical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.H11: information on b5 corrected.
|
|
Event Description
|
It was reported that the motor was not working.No case involved.Results of investigation have concluded that this unit had a blade stall which makes it a reportable event.
|
|
Search Alerts/Recalls
|