H3: product analysis: the inner hub spun easily by hand however the tip was broken off which would have resulted in the reported event.The detached portion of the distal inner tip measured 0.68¿.Spiral wrap breakage may be caused by unusual torsional loads.The distal inside diameter of the outer tube was rough and worn which is typical of friction caused by excess pressure during use or excess speed.The adjacent inner assembly was rough and worn.There was no allegation of damage to the bur prior to use.The information indicates excess pressure was applied while in the handpiece during rotation which caused friction and excess torsional loads resulting in the tip breaking off at the spiral wrap.H6: the fdm 4114, fdr 3221 and fdc 67 are no longer applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|