• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST MESH; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST MESH; SURGICAL MESH Back to Search Results
Catalog Number 5954124
Medical Device Problem Codes Defective Device (2588); Patient-Device Incompatibility (2682); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Abdominal Pain (1685); Adhesion(s) (1695); Necrosis (1971); Injury (2348); Obstruction/Occlusion (2422)
Date of Event 07/25/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventralight st on (b)(6) 2019.As reported, the patient is making a claim for an adverse patient outcome against the ventralight st.It is alleged that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-dec-2018) is considered to be a best estimate.Updated fields: a2, b3, b4, b5, b6, b7, d4 (product catalog no., corporate lot no., udi no., expiry date), g3, g6, h2, h4, h6, h10.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventralight st on (b)(6) 2019.As reported, the patient is making a claim for an adverse patient outcome against the ventralight st.It is alleged that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2019 ¿ patient was diagnosed with ventral abdominal hernia thereby underwent open repair with implant of ventralight st mesh.Per operative notes, ¿intraabdominal adhesions and hernia sac in the subcutaneous tissues above the fascia was excised.The ventralight st mesh was placed over the defect and sutured.¿ (b)(6) 2019 - patient was diagnosed with closed loop bowel obstruction with abdominal pain thereby underwent open repair.Per operative notes, ¿adhesions were removed.There were dilated loops of small bowel and soft tissue mainly fat that seemed necrotic and non-viable were noted.The small bowel was decompressed and serosa was closed with sutures and wound vac was placed.¿ (b)(6) 2019 ¿ patient was diagnosed with dilated small bowel thereby underwent fascial closure with internal retention sutures.Per operative notes, ¿the fascia was closed with sutures.The wound was irrigated and wound vac was place.¿ (b)(6) 2020 - patient was diagnosed with dilated small bowel thereby underwent debridement of midline abdominal wound.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key10216472
Report Number1213643-2020-06244
Device Sequence Number19500295
Product Code FTL
UDI-Device Identifier00801741031601
UDI-Public(01)00801741031601(17)210128(10)HUCZ2150
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K101851
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/28/2021
Device Catalogue Number5954124
Device Lot NumberHUCZ2150
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2020
Supplement Date Received by Manufacturer07/14/2025
Initial Report FDA Received Date07/01/2020
Supplement Report FDA Received Date08/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/15/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
-
-