WARSAW ORTHOPEDICS CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 5442147
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent thoracolumbar/lumbar spine surgery due to burst fracture of the lumbar 4 vertebral body.Intra-op, while using crosslink for breaking screw plug, the lower side of one side of the screw plug broke at the threaded joint.It was removed intra-operatively.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
Additional information: d9, h3, h6 h3.Device evaluation summary: visual and optical inspection did not reveal any cracks or deformation to the crosslink.No signs of inner or outer thread damage.Both of the set screws have been damaged.The lower cone shaped portion of the set screws has been sheared off.This type of damage is consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key10216723
Report Number1030489-2020-00764
Device Sequence Number19434183
Product Code KWP
UDI-Device Identifier00613994588487
UDI-Public00613994588487
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K091974
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5442147
Device Catalogue Number5442147
Device Lot Number0731866W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2020
Supplement Date Received by Manufacturer11/06/2020
Initial Report FDA Received Date07/01/2020
Supplement Report FDA Received Date11/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
Patient Weight58
-
-