| Model Number |
MMT-523NAP |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the buttons on the insulin pump were unresponsive at times.Customer reported buttons were sticky and hard to press.Customer also reported that battery life was diminished.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to partial unlocked lcd keypad connector noted during visually inspect.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test or verify charge or battery lasts less than expected anomaly due to buttons not responding.(b)(4).
|
| |
|