It was reported that the blood draw was being done using an abs-10010s, acp double syringe system.The kit exploded over the doctor, patient and the room.It was cleaned up and the doctor repeated the process.(b)(6) year old patient.Additional information obtained 6/18/2020: when the surgeon was conducting the procedure the inner syringe plunger was pushed out from the pressure inside and the patient's blood went all over the room.It was on the floor, walls, the surgeon's clothing, shoes, and on the patient as well.The office had to complete the procedure using another kit.The room was then closed for the remainder of the day while the blood spill was cleaned per blood borne pathogen cleaning process.The device was discarded due to blood exposure.
|