| Model Number |
10604 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.The target lesion was located in the right coronary artery.A 4.00 x 24 synergy ii drug-eluting stent was advanced for treatment.The stent was in and out of the guide catheter five times; however, during the last time, the stent was caught in the guide or tuohy borst.Upon removal, it was noticed that the stent struts at the proximal end of the stent were bent.The procedure was completed with another synergy stent.No patient complications were reported and the patient's status was stable.
|
| |
|