NIHON KOHDEN CORPORATION ZM-521PA; TRANSMITTER Back to Search Results
Model Number ZM-521PA
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that their transmitter intermittently shuts off and on.The customer also reported that when they connect this device to new batteries, it takes a while for it to start up.No harm or injury was reported.They will be sending this transmitter in for an exchange.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event or Problem Description
The customer reported that their transmitter intermittently shuts off and on.
 
Event or Problem Description
The customer reported that their transmitter intermittently shuts off and on.
 
Additional Manufacturer Narrative
Details of complaint: the customer reported that the transmitter would intermittently shut off and then power back on randomly.They also reported that when replacing the batteries with new ones, it would take a while for it to start up.No harm or injury was reported.Investigation summary: startup failures attributed to the zm 5xxx transmitters may be caused by physical damage, battery failure or power board failure.A review of the history of the serial number identified no other reports of the issue.Based on the available information a definitive root cause could not be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-521PA
Common Device Name
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key10219656
Report Number8030229-2020-00351
Device Sequence Number14054608
Product Code DRT
UDI-Device Identifier04931921115084
UDI-Public4931921115084
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K043517
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
Report Date (Section F) 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZM-521PA
Device Catalogue NumberZM-521PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/11/2020
Device Age92 MO
Event Location Hospital
Date Report to Manufacturer07/01/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2020
Supplement Date Received by Manufacturer02/01/2023
Initial Report FDA Received Date07/01/2020
Supplement Report FDA Received Date02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/24/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA07/01/2020
-
-