| Model Number |
ZM-521PA |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer reported that their transmitter intermittently shuts off and on.The customer also reported that when they connect this device to new batteries, it takes a while for it to start up.No harm or injury was reported.They will be sending this transmitter in for an exchange.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that their transmitter intermittently shuts off and on.
|
| |
|
Event or Problem Description
|
|
The customer reported that their transmitter intermittently shuts off and on.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint: the customer reported that the transmitter would intermittently shut off and then power back on randomly.They also reported that when replacing the batteries with new ones, it would take a while for it to start up.No harm or injury was reported.Investigation summary: startup failures attributed to the zm 5xxx transmitters may be caused by physical damage, battery failure or power board failure.A review of the history of the serial number identified no other reports of the issue.Based on the available information a definitive root cause could not be identified.
|
| |
|
Search Alerts/Recalls
|