| Model Number |
367008 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A pantera leo coronary balloon system was selected for treatment of a severely calcified lesion.The balloon ruptured at 12 atm during inflation.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned delivery system was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed during functional testing a fine jet of water distal to the distal x-ray marker.The microscopic analysis of the balloon surface showed a microscopic pinhole and scratches on the balloon surface nearby the damage site review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The damage of the balloon surface is most likely related to the patients anatomy.
|
| |
|