Model Number 459888 |
Device Problems
Pacing Problem (1439); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
|
Patient Problems
Muscle Stimulation (1412); Undesired Nerve Stimulation (1980); Twiddlers Syndrome (2114)
|
Event Date 03/24/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id: 6984m adaptor, implanted: (b)(6) 2015.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient complained of phrenic nerve stimulation and diaphragmatic stimulation.The left ventricular (lv) lead was turned off due to an apparent gross dislodgement and high threshold.A lead revision was planned for a later date.No further patient complications have been reported as a result of this event.
|
|
Event Description
|
It was further reported that there was a loss of pacing due to the lv lead dislodgement.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Product event summary: the full lead was returned and analyzed.No anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that during the scheduled lv lead revision, the lv lead was noted to have pulled back entirely into the device pocket.The clinician suspected twiddler's syndrome and the lv lead was explanted and replaced.
|
|