• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ENSITE VELOCITY NAVLINK MODULE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ENSITE VELOCITY NAVLINK MODULE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H420995
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/09/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission. .
 
Event or Problem Description
During an atrial fibrillation and atrioventricular nodal re-entrant tachycardia (avnrt) ablation procedure, the navlink pin detached within the patch connector and mapping was no longer possible.The avnrt ablation was completed without mapping, however the atrial fibrillation portion of the procedure was unable to be completed and the procedure was cancelled.There were no adverse consequences to the patient due to the cancellation.
 
Additional Manufacturer Narrative
Additional information: d10, g4, g7, h2, h3, h6, h10.One ensite velocity¿ navlink module was returned for investigation.Inspection of the input and output connectors revealed the ¿neck¿ patch electrical contact pin has broken off and was not returned.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information provided to abbott and the investigation performed, the root cause of the event was isolated to a broken electrical contact pin.The cause of the detached pin remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY NAVLINK MODULE
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key10220707
Report Number2184149-2020-00094
Device Sequence Number19434044
Product Code DQK
UDI-Device Identifier05414734209953
UDI-Public05414734209953
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K160187
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberH420995
Device Catalogue NumberH420995
Device Lot Number263160004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/09/2020
Supplement Date Received by Manufacturer07/20/2020
Initial Report FDA Received Date07/01/2020
Supplement Report FDA Received Date07/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-