| Model Number |
00884450026571 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a peripheral vascular intervention procedure, the catheter tip detached within the patient.The physician had successfully acquired left retrograde femoral artery access with the 4f 90cm catheter and was attempting to negotiate the patient's descending aorta under fluoroscopy when the catheter tip detached.The physician successfully removed the foreign body from the patient's descending aorta with a vascular snare device.The patient tolerated the procedure well.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A review of the device history and complaint database could not be performed since the lot number was not provided.Should the device be returned at a later time, the investigation will be re-opened.
|
| |
|