Catalog Number SPL12506X
Device Problems Deflation Problem (1149); Inflation Problem (1310)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/29/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
During a procedure an attempt was made to use one sprinter legend rx ptca balloon catheter to treat a severely tortuous, severely calcified lesion exhibiting cto (chronic total occlusion-100%) located in the distal circumflex (cx) artery.There was no issues noted to the packaging.There were no issues noted when removing the device from the hoop.The device was not inspected.Negative prep was not performed.The lesion was not pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device.Excessive force was not used.It was reported that balloon deflation difficulties occurred and the device would not deflate at the lesion site.The device was removed from the patient by slowly withdrawing the balloon.The patient was reported to be alive with no injury.
 
Manufacturer Narrative
Product analysis summary: the stylette was partially loaded in the device and could not be removed.The strain relief was deformed.The transition shaft was stretched and flattened.The distal shaft was stretched.There was no damage to the proximal balloon bond.The balloon folds were intact; inflation had not been performed.The distal tip was necked.It was not possible to inflate the balloon in order to perform deflation testing due to the condition of the transition shaft and distal shaft.There was no other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: 8 atm of pressure was applied during inflation.The device was not moved or re-positioned in the lesion while inflated.It was stated that difficulties were noted during inflation of the device.It was stated that there was partial inflation of the balloon.It was also reported that balloon deflation difficulties occurred after the first inflation the balloon deflated slowly.The delivery balloon was removed normally after deflating the balloon with no difficulties noted.The same inflation device was used successfully with other devices.The concentration of contrast / saline used was 200mg/ml.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Correction: initial reporter phone number.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINTER LEGEND RX
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
MDR Report Key10223096
MDR Text Key197595135
Report Number9612164-2020-02390
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
PMA/PMN Number
K103095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 09/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/16/2022
Device Catalogue NumberSPL12506X
Device Lot Number218098165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2020
Date Manufacturer Received09/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age57 YR
Patient Weight50
-
-