| Model Number |
468-450 |
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
05/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation found no instrument marks or damage anywhere on the implant other than the implant being bent.The implant was activated with heat, and the implant hook closed properly and the bent shaft straightened as designed.The implant performed as intended.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon placed the prosthesis successfully several months ago.At least weeks/months later at follow up appointment, patient had mentioned to the surgeon that their hearing had not improved.After months of no improvement the physician decided to go in and investigate.When the physician opened up the patient, he found the shaft bent.The surgeon decided to place another prosthesis to restore the patient's hearing.
|
| |
|