Model Number N65-1
Medical Device Problem Codes Image Display Error/Artifact (1304); Temperature Problem (3022)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device had an error code 279 and a smoke was observed.One of the spring was loosen and shorted to the main board.They need to manually turn off the device.No flame but a little bit of smoke was confirmed.There was no patient involvement reported.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the spo2 board was defective.It was reported that there were signs of a thermal issue and the product gave an error message of error code 276.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KR  00000
MDR Report Key10227695
Report Number2936999-2020-00462
Device Sequence Number515112
Product Code DQA
UDI-Device Identifier10884521096738
UDI-Public10884521096738
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K051352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN65-1
Device Catalogue NumberN65-1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/16/2020
Supplement Date Received by Manufacturer08/20/2020
Initial Report FDA Received Date07/03/2020
Supplement Report FDA Received Date09/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-