Merge unity pacs is a medical image and information management system that is used for viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.On (b)(6) 2020, the customer reported receiving an exam report with no report body to (b)(6) health imaging on the same day.Troubleshooting from (b)(6) health imaging support found that there were no patient demographics, body, or conclusion data included in the report file.There was a report save event in the audit trail.Using the revision button from the dictation dialog box, the report template was re-generated and the report was then viewable as expected.This has the potential to delay patient treatment and/or diagnosis.There have been no reports of patient injury or harm as a result of this issue.Reference complaint (b)(4).
|