BOSTON SCIENTIFIC CORPORATION VISIONIST CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U225
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Device Sensing Problem (2917)
Patient Problem No Code Available (3191)
Event Date 04/04/2020
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.(b)(4).
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) detected high out-of-range right atrial (ra) lead impedance measurements resulting in a lead safety switch from bipolar to unipolar.The unipolar ra pacing impedance measurement was within normal limits.A review of stored episodes noted an isolated incident of oversensed noise however no noise could be produced with in clinic provocations.The clinician noted that the right ventricular (rv) lead threshold measurements had increased so the pulse width was changed.No adverse patient effects were reported.The device remains implanted and in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key10227950
MDR Text Key197597175
Report Number2124215-2020-11808
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559426
UDI-Public00802526559426
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/24/2017
Device Model NumberU225
Device Catalogue NumberU225
Device Lot Number700844
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/24/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age32 YR
-
-