| Catalog Number |
105762 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an external fluid leak was observed during patient treatment from a crack of the luer connector of the accessory line gmb-sp127 - y connector.A ¿flow problem¿ alarm was triggered by the prismaflex machine.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d4: lot #: the involved lot number was not confirmed.It was reported lot #s 18k097331 and 19j28t028 were in use at the time of the reported event.Lot # 18k09t331 was manufactured on nov 17 2018.Lot # 19j28t028 was manufactured on oct 08 2019.H10: the actual device was not available; however, a retained sample was provided for evaluation.Visual inspected did not reveal any obvious damage.The retention samples were also gravity tested and leak tested under water and no leak was observed.The reported condition was not verified.A batch review of both lots 18k097331 and 19j28t028 was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|