• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE LIFEBAND; CARDIAC RESUSCITATOR BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE LIFEBAND; CARDIAC RESUSCITATOR BAND Back to Search Results
Model Number MODEL 100
Medical Device Problem Codes Material Integrity Problem (2978); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll has not received the product for investigation.A follow-up report will be submitted when the product is returned and investigation has been completed.The death was not related to the autopulse device.The autopulse is used as an adjunct to manual cpr in cases of clinical death.The benefit of using the autopulse is that it in part substitutes mechanical compressions for the physical labor of manual chest compressions.If the autopulse did not start or unexpectedly stops compressions, rescuer should revert to manual cpr, which is the standard of care.In this case, during the autopulse platform (sn: (b)(4)) was used, it displayed an error message on the screen panel.The crew reposition the patient on multiple occasions and replaced the autopulse li-ion battery but was unable to clear the advisory.Crew immediately performed manual cpr.Also, the belt clip on the lifeband (lot #97059) was broken during patient use.Patient death was reported.The autopulse was intended to be used as an adjunct to manual cpr on adult patients.In case of stoppage of autopulse the trained user reverts to manual cpr.The transition from autopulse to manual cpr by trained users is similar to the time necessary for rescuer rotation, and presents the same workflow as manual cpr.Hence, based on available information, the patients' outcome was not negatively impacted by the interruptions when compared to standard of care manual cpr.Out-of-hospital cardiac arrest (ohca) is one of the main causes of death in industrial nations.About 25% of patients survive this event and make it to the hospital, and even fewer patients survive after 24 hours (nichol, nejm, 2015).In the united states, survival to hospital discharge after non-traumatic emergency medical services-treated cardiac arrest with any first recorded rhythm was 10.6% for patients of any age.Of the bystander-witnessed out-of-hospital cardiac arrests in 2011, 31.4% of victims survived to hospital discharge (mozaffarian, circulation, 2016).Death is an expected outcome for ohca.
 
Event or Problem Description
During patient use, customer reported that the autopulse platform (serial #(b)(4)) displayed user advisory - "(ua) 07" (discrepancy between load 1 and load 2 too large) on the screen panel.The crew reposition the patient on multiple occasions and replaced the autopulse li-ion battery but was unable to clear the advisory.Crew immediately performed manual cpr.Also, the belt clip on the lifeband (lot #97059) was broken during patient use.Patient death was reported.The autopulse platform associated with this lifeband is submitted under mfr 3010617000-2020-00640.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE LIFEBAND
Common Device Name
CARDIAC RESUSCITATOR BAND
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
MDR Report Key10228496
Report Number3010617000-2020-00641
Device Sequence Number14251814
Product Code DRM
UDI-Device Identifier00849111001588
UDI-Public00849111001588
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0701-08
Device Lot Number97059
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2020
Initial Report FDA Received Date07/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-