STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Model Number |
6020-0230 |
| Medical Device Problem Code |
Material Erosion (1214)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
06/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Typical cone corrusion with v40 cone and metal head size 36 according to known problems in the literature.Presence of a massive pseudotumour including the artificial hip joint from the os ilium to the trochanter minor of the femur, as well as via the foramen ischiadicum into the small pelvis with an extension of approx.5x5 cm according to mri.Necessary revision with pseudotumour removal and head exchange (special adapter).
|
| |
|
Search Alerts/Recalls
|
|
|