| Model Number |
MMT-723NAS |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic a that the insulin pump had motor error alarm.The customer stated that the drive support cap was normal.No harm a medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test, rewind test, basic occlusion test, prime or a33 test, excessive no delivery test and occlusion test, no motor error alarm during testing noted.The motor was tested outside the device and passed.(b)(4).
|
| |
|