OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-2T160
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device has not been returned to omsc but was returned to olympus (b)(4) (oekg).Oekg sent the device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the sample collected from the device.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event or Problem Description
Olympus medical systems corp.(omsc) was informed that as a result of microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device.Instrument channel: fungal hyphae (>20 cfu/endoscope), bacillus species (>20 cfu/endoscope), coagulase-negative staphylococci (>20 cfu/endoscope) the device had been reprocessed with an olympus automated endoscope reprocessor model etd3 (not available in the usa), using glutaraldehyde.There was no report of infection associated with this report.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10230998
Report Number8010047-2020-03883
Device Sequence Number2655498
Product Code FDS
UDI-Device Identifier04953170307492
UDI-Public04953170307492
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K954451
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberGIF-2T160
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/10/2020
Supplement Date Received by Manufacturer07/29/2020
Initial Report FDA Received Date07/05/2020
Supplement Report FDA Received Date08/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-