• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G29983
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Patient Involvement (2645)
Event Date 06/23/2020
Event Type  malfunction  
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Occupation: operation room team lead.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event Description
As reported, a flexor ansel guiding sheath's outer package was not sealed.The components fell out of the outer package as the device was hung for storage.The individual components inside the package seemed "fine".The device was not used.
 
Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Additional information was received.The bottom outer package was not sealed.Reportedly, the content inside was "fine", so the device was used.
 
Event Description
No new patient or event information to report.
 
Manufacturer Narrative
Description of event: as reported, a flexor ansel guiding sheath's outer package was not sealed.The components fell out of the outer package as the device was hung for storage.The individual components inside the package seemed "fine", so the device was used.Investigation ¿ evaluation.A document based investigation was also performed including a review of complaint history, device history record, documentation, manufacturing instructions, and quality control data.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was made out of specification.A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of complaint history records shows no other complaints associated with the complaint device lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: how supplied supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.A capa was previously opened in regard to missing bottom seal and this capa was closed to voe (verification of effectiveness) as it was found, that during the sealing process an operator can have multiple unsealed product in their hands and then inadvertently not perform the sealing operation on all pouches being held.This can lead to the seal being missed and mixed in with sealed product.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Using this information we were able to determine that the definitive root cause of the failure mode was related to the manufacturing and packaging of the device.Per the quality engineering risk assessment, no further action is warranted.Cook will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Type of Device
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10232838
MDR Text Key197598002
Report Number1820334-2020-01250
Device Sequence Number1
Product Code DYB
UDI-Device Identifier00827002299832
UDI-Public(01)00827002299832(17)221021(10)10095124
Combination Product (y/n)N
PMA/PMN Number
K142829
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/21/2022
Device Model NumberG29983
Device Catalogue NumberKCFW-6.0-18/38-45-RB-ANL2-HC
Device Lot Number10095124
Was Device Available for Evaluation? No
Date Manufacturer Received12/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-