• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM LCP DISTAL TIBIA T-PLATE 3 HOLE/64MM; PLATE,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM LCP DISTAL TIBIA T-PLATE 3 HOLE/64MM; PLATE,FIXATION,BONE Back to Search Results
Model Number 02.112.203
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Ulcer (2274); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent a hardware removal of the locking compression plating system (lcp) of the distal tibia due to infection.Initially, the patient was implanted with an lcp system on (b)(6) 2018.The surgeon commented that the patient had developed skin ulcers early on, which caused infection from wearing his post-op boot too tight.The hardware was removed, and the removal surgery was completed.The patient is ok.Concomitant devices reported: lcp one-third tubular plate with collar 5 holes/57mm (part number 241.351, lot unknown, quantity 1); 2.7/3.5mm va-lcp anterolateral distal tibia pl/6 holes/left (part number 02.118.205, lot unknown, quantity 1); screws: locking (part number unknown, lot unknown, quantity 11); screws: cortex (part number unknown, lot unknown quantity 9).This product complaint, (b)(4), is related to (b)(4) and (b)(4).(b)(4)captures the 10 of 23 devices, (b)(4) captures the 10 of 23 devices, and (b)(4) captures the 3 of 23 devices.This report involves one (1) 3.5mm lcp distal tibia t-plate 3 hole/64mm.This is report 2 of 10 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5MM LCP DISTAL TIBIA T-PLATE 3 HOLE/64MM
Common Device Name
PLATE,FIXATION,BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10233231
Report Number2939274-2020-03078
Device Sequence Number2426456
Product Code HRS
UDI-Device Identifier10886982034373
UDI-Public(01)10886982034373
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K080522
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.112.203
Device Catalogue Number02.112.203
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2020
Initial Report FDA Received Date07/06/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-