• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Model Number G52331
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Occupation: director.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, prior to use during an angiogram, the seal was noted to be missing from an advance 35 lp low profile balloon catheter package.As the user opened the package, the seal was noted to be missing and the balloon was missing from the shaft.The device did not make contact with any patient.A new device of the same type was used to complete the procedure.
 
Event or Problem Description
No new patient or event information to report.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Description of event: as reported, prior to use during an angiogram, the seal was noted to be missing from an advance 35 lp low profile balloon catheter package.As the user opened the package, the seal was noted to be missing and the balloon was missing from the shaft.Investigation evaluation: a visual inspection and functional testing of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record, drawings, the instructions for use, manufacturing instructions, quality control data, and specifications.One advance 35 lp low profile balloon catheter was received.The device was returned prior to use and the package was opened on the top and side seal.There was evidence of the seal present, it appeared the package had been opened after it was sealed.The balloon catheter was placed in spiral holder but was incorrectly positioned.Upon removal and inspection it was discovered that the balloon portion of the catheter was missing.The catheter length was 76.5cm.No further issues were noted during the analysis.A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of complaint history records shows no other complaints associated with the complaint device lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: how supplied supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Based on the investigation results it was most likely that the packaging was inadvertently damaged during either shipping or storage.There were no issues noted during the manufacturing review to suggest that manufacturing issues caused this failure.Per the quality engineering risk assessment, no further action is warranted.Cook will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10233454
Report Number1820334-2020-01255
Device Sequence Number13601804
Product Code DQY
UDI-Device Identifier10827002523316
UDI-Public(01)10827002523316(17)220906(10)9996240
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K091527
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2022
Device Model NumberG52331
Device Catalogue NumberPTA5-35-80-4-2.0
Device Lot Number9996240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2020
Initial Date Received by Manufacturer 06/26/2020
Supplement Date Received by Manufacturer12/08/2020
Initial Report FDA Received Date07/06/2020
Supplement Report FDA Received Date12/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-