| Model Number |
CL-841 |
| Medical Device Problem Codes |
Product Quality Problem (1506); Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Tissue Damage (2104); No Code Available (3191); Unspecified Tissue Injury (4559)
|
| Date of Event |
06/18/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a caesarean section, when they closed the uterus, the device made the wound bleed of not more than 500 cc.They used another suture to sew the wound.An additional operation was performed to correct the issue.There was tissue damage.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a caesarean section, when the user closed the uterus, the device made the wound bleed of not more than 500cc but did not break.The needle was bigger than the thread, the thread was strong but difficult to absorb tissue fluid.The operator used another suture from a different manufacturer to sew the wound.There was tissue damage.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.One sealed representative sample with the appropriate pack aging was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the needle did not appear as expected.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are tho roughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: absorption of polysorb tm sutures is essentially complete between the 56th and 70th day.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|