• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION KNEE; INSERT, STD., PRIMARY, SIZE6 9MM CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION KNEE; INSERT, STD., PRIMARY, SIZE6 9MM CM Back to Search Results
Model Number 385-09-506
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to these documents as soon as it becomes available.
 
Event or Problem Description
Revision surgery - patient had knee instability.Surgeon determined it was due to poly wear.Scheduled a poly swap.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported instability.The actual length of in-vivo for the item listed is unknown, without an invoice/dticket the original surgery date could not be established.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material report (ncmr) associated with the products that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to instability.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION KNEE
Common Device Name
INSERT, STD., PRIMARY, SIZE6 9MM CM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx
MDR Report Key10234329
Report Number1644408-2020-00549
Device Sequence Number594817
Product Code JWH
UDI-Device Identifier00888912117746
UDI-Public(01)00888912117746
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K000590
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/28/2013
Device Model Number385-09-506
Device Catalogue Number385-09-506
Device Lot Number53942649
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/08/2020
Supplement Date Received by Manufacturer07/07/2020
Initial Report FDA Received Date07/06/2020
Supplement Report FDA Received Date08/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-