MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. TURON SHOULDER; HEAD, HUMERAL, NEUTRAL, 46-16

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. TURON SHOULDER; HEAD, HUMERAL, NEUTRAL, 46-16 Back to Search Results
Model Number 520-46-016
Health Effect - Clinical Codes Failure of Implant (1924); No Code Available (3191)
Date of Event 06/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as glenoid wear.The previous surgery and the surgery detailed in this event occurred 6 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to glenoid wear.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient developed significant glenoid wear, and needed to be resurfaced.The parts listed were removed to allow access to the glenoid since the primary surgery was a hemi arthroplasty.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
HEAD, HUMERAL, NEUTRAL, 46-16
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin texas 78758-5445,
MDR Report Key10234339
Report Number1644408-2020-00525
Device Sequence Number594818
Product Code KWS
UDI-Device Identifier00888912145145
UDI-Public(01)00888912145145
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2018
Device Model Number520-46-016
Device Catalogue Number520-46-016
Device Lot Number917C1011
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2020
Initial Report FDA Received Date07/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000 LOT 878C1345
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
-
-