| Model Number |
B1040-020 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other similar complaints.Based on the information provided, the difficulty to advance and subsequent balloon rupture appear to be due to case circumstances.It is likely that the device encountered resistance due to interaction with the heavily calcified and heavily tortuous lesion causing damage to the outer surface of the balloon material and resulted in the rupture.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a heavily calcified lesion in the heavily tortuous, concentric left superficial femoral artery (sfa).The 4.0x20 mm armada 35 percutaneous transluminal angioplasty (pta) catheter was advanced with noted resistance with the calcified lesion.The armada 35 balloon was inflated to nominal pressure and ruptured.A new armada pta catheter was used to complete the procedure.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
|
| |
|