Model Number 72404281
Medical Device Problem Codes Collapse (1099); Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 06/11/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had a pump replacement surgery due to a dimpled pump with his inflatable penile prosthesis (ipp).
 
Additional Manufacturer Narrative
Correction to h2, h3, and h6: updated to include device analysis.Product analysis summary product analysis confirmed the reported event related to pump collapse and mechanical issue.The ams 700 momentary squeeze (ms) pump was visually inspected.The kink resistant tubing (krt) was worn to the filament; however, no leaks were present.The pump was functionally tested and failed activation test.Product analysis concluded that identified a pump failed activation test was the most probable cause of the device malfunction.Dhr review / similar complaint review of manufacturing documentation was performed to ensure that all required in-process and final inspections and testing were completed.Review of the manufacturing records for batch 1000002554 and it respective container 21289053, found no evidence that the device failed to meet applicable product specifications prior to shipment from boston scientific.There is no evidence that the reported events are related to a manufacturing deficiency based on the product analysis and a review of the manufacturing records.Additionally, potential emerging trends are captured as part of the post market signal evaluation and escalation process.Risk review the ams 700 hazard analysis indicates that the as reported events of this complaint do not represent a new hazardous situation.Additionally, boston scientific does not have responsibility to provide training to medical professional as documented in the dfu/ifu, and the ams 700 is not a new product because it is not within the first six months of its release in any geography, and it is not a new design platform.No further review is required.
 
Event or Problem Description
It was reported that the patient had a pump replacement surgery due to a dimpled pump with his inflatable penile prosthesis (ipp).It was added that the pump did not work and was "freezing" when trying to use it.The patient was experiencing these problems with the pump and also pain prompting him to see his physician.The onset of these issues was (b)(6)2020.The patient outcome following this surgery was "not bad.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Common Device Name
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
MDR Report Key10236434
Report Number2183959-2020-02814
Device Sequence Number3093341
Product Code FHW
UDI-Device Identifier00878953003849
UDI-Public00878953003849
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2022
Device Model Number72404281
Device Catalogue Number72404281
Device Lot Number1000002554
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2020
Initial Date Received by Manufacturer 06/11/2020
Supplement Date Received by Manufacturer07/19/2020
Initial Report FDA Received Date07/06/2020
Supplement Report FDA Received Date07/22/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
-
-