It was reported that during a knee scope procedure, the camera head presented lot of brightness.Backup device was available to complete the procedure.Delay greater than 30 minutes was reported.However, no patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H3, h6: the reported device, used in treatment, was returned to the designated complaint station for independent evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed and no deficiencies were observed.A functional test revealed that the displayed image is bright and does not respond to the electronic level control function.The complaint was confirmed and the root cause has been associated with a component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures.Factors, unrelated to the manufacturing and design of the device that could have contributed to the damage, include a failing cable or sensor component.No containment or corrective actions are recommended at this time.
|