|
The device, used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is chipped and damaged and has signs of wear and tear from use.The device was manufactured in 2015.According to clinical medical investigation , currently unable to rule out a procedural variance as a contributing factor to the reported event, which does not represent a device malfunction.Based on the information provided, there was no surgical delay or patient injury/impact as the procedure was completed with the same device; therefore, no further medical assessment is warranted at this time.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaints.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.Credit will be issued for the device.
|