U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. IPC HANDPIECE - XPS STRAIGHTSHOT M4; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. IPC HANDPIECE - XPS STRAIGHTSHOT M4; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE Back to Search Results
Model Number 1898200TRF
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider reported that the handpiece was overheating during the procedure.There was no patient impact.
 
Additional Manufacturer Narrative
H3: analysis could not verify the reported complaint.The unit came in with seized thumbwheel and corrosion on motor/cable assembly.The motor was corroded that the handpiece could not run that is why it could not be tested.The unit was evaluated and scrapped.It was beyond repair.H6: the fdm 4118, fdr 3233 and fdc 11 are no longer applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC HANDPIECE - XPS STRAIGHTSHOT M4
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key10238896
Report Number1045254-2020-00301
Device Sequence Number2864719
Product Code ERL
UDI-Device Identifier00613994598301
UDI-Public00613994598301
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K041413
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1898200TRF
Device Catalogue Number1898200TRF
Device Lot Number206378933
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2020
Supplement Date Received by Manufacturer07/15/2020
Initial Report FDA Received Date07/07/2020
Supplement Report FDA Received Date08/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-