Catalog Number 8606000
Device Problems Gas Output Problem (1266); Failure to Deliver (2338); Device Sensing Problem (2917)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/18/2020
Event Type  malfunction  
Manufacturer Narrative
Log file evaluation gave an indication that a problem with the motor position detection system in the ventilator unit was the root cause for the ventilator failure.Further examination of the device hardware confirmed this; the light barrier and the dedicated cable have been replaced which has fully solved the device problem.Dräger finally concludes that the device responded as designed upon the malfunction of a single electronic subsystem which has failed after approximately 12 years of operation.The issue is rated to be an isolated case.No patient consequences have occurred.
 
Event Description
It was reported that the device posted a ventilator failure alarm during use.There was no injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FABIUS TIRO
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
MDR Report Key10238905
MDR Text Key198542645
Report Number9611500-2020-00228
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042419
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial
Report Date 07/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8606000
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-