TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL PUMP ADAPTER Back to Search Results
Model Number 824043
Medical Device Problem Code Noise, Audible (3273)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, the sarns delphin pump made an unusual grinding noise.As per the user facility, they heard an unusual grinding noise emanating from the sarns delphin pump or pump adapter and believed that the noise was related to the adapter.The pump was primed the day before and was left on the heart lung machine.The chief of perfusion opted to use the circuit as in on case that morning and did not change the delphin pump or adapter.They did noticed a mild grinding sound from the pump but adjusted or turned the delphin head in the adapter and the noise subsided.No known impact or consequence to patient.The product was not changed out.Procedure completed successfully.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on july 7, 2020.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information).H4 (device manufacture date).H6 (identification of evaluation codes 11, 3331, 4114, 170, 23).Method code #1: 11 - testing of device from same lot/batch retained by manufacturer.Method code #2: 3331 - analysis of production records.Method code #3: 4114 - device not returned.Results code: 170 - manufacturing process problem identified.Conclusions code: 23 - manufacturing deficiency.The complaint sample was not returned for evaluation, so a thorough investigation could not be performed.Retention samples are not maintained for this product code; therefore, a pump retention sample was used.The pump retention sample was inspected with no visual anomalies noted.The unit was setup in a saline circuit with a sarn drive system and a sorin drive system with a terumo cp adapter.While pumping with the sarns system and the sorin drive, the pump exhibited no unusual noises.Research of pump production records showed that mold maintenance was performed to improve the dimensional tolerance of the molded components and associated bearings.This maintenance was performed after component production for the affected sample.While the mold was not out of specification, this preventive maintenance and mold modification improved the bearing/housing interface with improved tolerances.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: e1 (address corrected establishment name).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information).Based on this investigation, it is believed that the root cause of this failure mode is a combination of a poor technique in using the arbor press for applying the mag rotor assembly into the separator and the dimension of the bearing pocket in the magnet housing.Corrections are being investigated as it relates to the arbor press to ensure consistent assembly of the bearing and the magnet rotor.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL PUMP ADAPTER
Common Device Name
CENTRIFUGAL PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key10239413
Report Number1124841-2020-00152
Device Sequence Number16920741
Product Code KFM
UDI-Device Identifier00699753450417
UDI-Public(01)00699753450417
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K112229
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model Number824043
Device Catalogue NumberN/A
Device Lot NumberRM22
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/18/2020
Supplement Date Received by Manufacturer08/18/2020
11/23/2020
Initial Report FDA Received Date07/07/2020
Supplement Report FDA Received Date08/20/2020
12/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-