WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11MM TI CANNULATED TIBIAL NAIL-EX/345MM-STERILE; NAIL, FIXATION, BONE Back to Search Results
Model Number 04.004.549S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Nonunion/Delayed-union Bone Fracture (2369); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/ investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent removal of the expert tibial nail, five (5) 5.0mm locking screws and two (2) 4.0mm cannulated screws.The reason for the removal of the implants was due to non-union and the patient anatomy fell into valgus from implant failure.During the revision procedure, all the original implants were removed and the patient was revised to a new tibial nail and fibula locking plate.The procedure was successfully completed without a surgical delay.Patient status is unknown.This report is for one (1) 11mm ti cannulated tibial nail-ex/ 345mm-sterile.This is report 1 of 8 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history: part number: 04.004.549s, 11mm ti cannulated tibial nail - ex/345mm - sterile.Lot number: 6456599 (sterile).Manufacturing location: (b)(4).Manufacturing date: 19-aug-2010.Expiration date: 31-jul-2019.Lot quantity: 6.Work order traveler met all inspection acceptance criteria.Inspection sheet, in-process acceptance sheet, met all inspection acceptance criteria.Inspection sheet, inspect dimensional, met all inspection acceptance criteria.Inspection sheet, final inspection, met all inspection acceptance criteria.Packaging label log, was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn 7039 supplied by sterigenics was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: part number: 21012, tialnbri13.00.Lot number: 6278837.Lot quantity: 2,902 lbs.Product traveler met all inspection acceptance criteria.Certified test report supplied by perryman company dated 10-apr-2009 and inspection certificate supplied to perryman by vsmpo dated 30-sep-2006 were reviewed and determined to be conforming.Raw material receiving/putaway checklist met all inspection acceptance criteria.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11MM TI CANNULATED TIBIAL NAIL-EX/345MM-STERILE
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10239605
Report Number2939274-2020-03121
Device Sequence Number2496322
Product Code JDS
UDI-Device Identifier10886982084491
UDI-Public10886982084491
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K040762
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Model Number04.004.549S
Device Catalogue Number04.004.549S
Device Lot Number6456599
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/23/2020
Supplement Date Received by Manufacturer08/04/2020
08/05/2020
Initial Report FDA Received Date07/07/2020
Supplement Report FDA Received Date08/05/2020
08/06/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-