• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 5.0MM TI LOCKING SCREW W/T25 STARDRIVE 34MM FOR IM NAILS; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 5.0MM TI LOCKING SCREW W/T25 STARDRIVE 34MM FOR IM NAILS; SCREW, FIXATION, BONE Back to Search Results
Model Number 04.005.524
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Nonunion/Delayed-union Bone Fracture (2369); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent removal of the expert tibial nail, five (5) 5.0 mm locking screws and two (2) 4.0 mm cannulated screws.The reason for the removal of the implants was due to non-union and the patient anatomy fell into valgus from implant failure.During the revision procedure, all the original implants were removed and the patient was revised to a new tibial nail and fibula locking plate.The procedure was successfully completed without a surgical delay.Patient status is unknown.This report is for one (1) 5.0 mm ti locking screw w/t25 stardrive 34 mm for im nails.This is report 6 of 8 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 corrected data: device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0MM TI LOCKING SCREW W/T25 STARDRIVE 34MM FOR IM NAILS
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10240267
Report Number2939274-2020-03130
Device Sequence Number2496327
Product Code HWC
UDI-Device Identifier10886982085924
UDI-Public10886982085924
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K000089
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.005.524
Device Catalogue Number04.005.524
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/23/2020
Supplement Date Received by Manufacturer08/05/2020
Initial Report FDA Received Date07/07/2020
Supplement Report FDA Received Date08/06/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-