| Model Number |
1219-28-054 |
| Medical Device Problem Codes |
Fracture (1260); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
05/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The description from the distributor states: on (b)(6) 2019 we helped the surgeon in a cementless thp.The surgery went very well.On saturday (b)(6) 2020, the surgeon contacted me to resume surgery for this patient.Wednesday (b)(6) 2020, (14 months later), we did the required surgery.Finding.The insert came out of its place and therefore became mobile.It is missing 3 notches out of 6.We looked for them, we didn't find them.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.
|
| |
|
Search Alerts/Recalls
|