| Model Number |
CD3257-40 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/16/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.Further information was requested but not received.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was prophylactically explanted and replaced due to the advisory.
|
| |
|
Event or Problem Description
|
|
New information received notes that the patient's condition was stable throughout the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5.
|
| |
|
Additional Manufacturer Narrative
|
|
Interrogation of the device revealed the device was at eri when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Based on this information, there was no evidence of premature depletion.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on 11 october 2016.
|
| |
|