| Model Number |
1012273-12 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that during unpacking of a 2.75x12mm trek balloon dilatation catheter (bdc), the connection piece of the hub was found to be broken off.The device was not used and there was no patient involvement.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported material separation hub was able to be confirmed; however, there was a noted misshaped hub.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined a conclusive cause for the reported/noted difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.Na.
|
| |
|