Model Number 97712
Device Problems Disconnection (1171); Improper or Incorrect Procedure or Method (2017); Battery Problem (2885)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 07/06/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer via a manufacturer representative (rep) regarding a patient with an implantable neurostimulator (ins).It was reported that the ins was overdischarged.The patient had not charged the battery in 3 months because she had a lot going on.The patient felt the leads came out because she felt a bump at the battery site.A physician recharge was unsuccessful on the first attempt.The patient was coming back in 3 weeks to try and jumpstart the battery again.If not, a battery replacement will be discussed.An x-ray was being scheduled to determine if the leads disconnected and the rep was going to follow up with the hcp regarding the x-ray results.No further complications were reported.
 
Manufacturer Narrative
Continuation of d11: product id 977a260, serial# (b)(6), implanted: (b)(6) 2016, product type lead.Product id 977a260, serial# (b)(6), implanted: (b)(6) 2016, product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the rep.It was reported that the results of the x-rays were not determined as patient cancelled her appointment.It was not determined if the leads were disconnected because the battery is dead.They could not jumpstart the battery.Patient will need a new battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10245320
MDR Text Key197964736
Report Number3004209178-2020-11770
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109506
UDI-Public00643169109506
Combination Product (y/n)N
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2017
Device Model Number97712
Device Catalogue Number97712
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/06/2020
Initial Date FDA Received07/08/2020
Supplement Dates Manufacturer Received08/25/2020
Supplement Dates FDA Received08/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Age76 YR
-
-