• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NEPHROMAX; CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NEPHROMAX; CATHETER, NEPHROSTOMY Back to Search Results
Model Number M0062101170
Medical Device Problem Codes Burst Container or Vessel (1074); Difficult to Remove (1528)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); No Known Impact Or Consequence To Patient (2692)
Date of Event 06/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a nephromax dilatation balloon catheter was used during a percutaneous nephrolithotomy (pcnl) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the balloon burst.Reportedly, the sheath got stuck in the tissue but was eventually taken out without extra bleeding.The procedure was completed a different device.No patient complications have been reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
 
Event or Problem Description
It was reported to boston scientific corporation that a nephromax dilatation balloon catheter was used during a percutaneous nephrolithotomy (pcnl) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the balloon burst.Reportedly, the sheath got stuck in the tissue but was eventually taken out without extra bleeding.The procedure was completed a different device.No patient complications have been reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.***additional information received on (b)(6), 2020*** it was reported that the anatomy location was in the kidney.Also, the balloon was noted to burst at under 20 atm.The patient had a minor bleeding as a result of this event; however, the patient was stable at the conclusion of the procedure.
 
Additional Manufacturer Narrative
Block h6: problem code 1074 captures the reportable event of balloon burst.Conclusion code 4316 is being used in lieu of an adequate conclusion code for device not returned.Block h10: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.Additional information: block b5 (describe event or problem) block e1 (physician's first name and initial reporter address 2) block h6 (patient code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPHROMAX
Common Device Name
CATHETER, NEPHROSTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10245885
Report Number3005099803-2020-02687
Device Sequence Number10896803
Product Code LJE
UDI-Device Identifier08714729012641
UDI-Public08714729012641
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K121614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/06/2022
Device Model NumberM0062101170
Device Catalogue Number210-117
Device Lot Number0023752593
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/17/2020
Supplement Date Received by Manufacturer07/07/2020
Initial Report FDA Received Date07/08/2020
Supplement Report FDA Received Date07/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-