WRIGHTS LANE SYNTHES USA PRODUCTS LLC 14MM CANNULATED DRILL BIT; BIT,DRILL Back to Search Results
Model Number 03.033.004
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is jnj representative.Investigation summary: investigation flow: damage visual inspection: the 14mm cannulated drill bit flexible (p/n 03.033.004, lot # 5l56841) was received at us cq broken on the laser cut teeth segment on the shaft.The shaft was completely broken into 2 pieces.No other issues were identified with the returned components of the device.Device failure/defect identified? yes.Dimensional inspection: dimensional inspection of the received device was performed.Document/specification review: based on the date of manufacture relevant drawings are reflecting the current and manufactured revision were reviewed.Complaint confirmed? yes, the device received was broken.Hence confirming the allegation.Conclusion: the complaint was confirmed for the 14mm cannulated drill bit flexible (p/n 03.033.004, lot # 5l56841) as the shaft of the device was completely broken.The unintended external forces might have contributed to the reported complaint condition.There was no indication that a design or manufacturing issue has caused the breakage and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: part number: 03.033.004, lot number: 5l56841, manufacturing site: (b)(4), release to warehouse date: aug.20, 2019.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance's were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the reporter found six (6) damaged items in sterile processing department (spd).There was no patient involvement.This complaint involves six (6) devices.This report is for (1) 14mm cannulated drill bit.This is report 3 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
14MM CANNULATED DRILL BIT
Common Device Name
BIT,DRILL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10247751
Report Number2939274-2020-03159
Device Sequence Number16942439
Product Code HTW
UDI-Device Identifier10886982271082
UDI-Public(01)10886982271082
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.033.004
Device Catalogue Number03.033.004
Device Lot Number5L56841
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2020
Initial Report FDA Received Date07/08/2020
Patient Sequence Number1
-
-