| Model Number |
24-010-14-71 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Impaired Healing (2378)
|
| Date of Event |
06/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Reference exemption number e2017029.Multiple mdr reports were filed for this event, please see associated report: mdr: 9610905-2020-00054; 9610905-2020-00056.
|
| |
|
Event or Problem Description
|
|
It was reported the product pulled from the bone and was removed due to patient non-compliance.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The failure root cause cannot be determined due to no device being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|
Search Alerts/Recalls
|