OBERDORF SYNTHES PRODUKTIONS GMBH 5.0MM TI LOCKING SCR SLF-TPNG W/T25 STARDRIVE 42MM-STERILE; APPLIANCE,FIXATION,NAIL
|
Back to Search Results |
|
| Model Number |
412.215S |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bone Fracture(s) (1870)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported additional fractures occurred post-operatively and were confirmed by x-ray for a femoral neck system (fns).Revision surgery occurred on (b)(6) 2020, in which the fns implants were removed and the patient revised to a bipolar hip arthroplasty (bha).The original open reduction internal fixation (orif) surgery for femoral neck fractures using the fns system occurred (b)(6) 2020.This is report 5 of 5 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative:the complaint condition that a fracture occurred right under the fns implants could be confirmed according to the received x-ray.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot part: 412.215s lot: 5l63290 manufacturing site: selzach; supplier: (b)(4); release to warehouse date: 06 aug 2019; expiry date: 01 july 2029; since there is no allegation against packing or sterility dhr review is done for non-sterile part: part number: 412.215; lot number: 5l20980; manufacturing site: mezzovico; release to warehouse date: 03 july 2019.A manufacturing record evaluation was performed for the non-sterile device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|