Model Number 1218-87-658
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 06/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the ceramax ceramic acetabular liner cracked in multiple pieces after impacting into acetabular shell.Left side.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4) investigation summary: examination of the returned liner confirms the reported material fracture.The root cause is attributed to inadvertent use error.No manufacturing or material problem identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: product code 121887658, work order 9268477 was manufactured on 11-sep-2019.20 parts were manufactured per specification and all raw materials met specification.Device history review: product code 121887658, work order 9268477 was manufactured on 11-sep-2019.20 parts were manufactured per specification and all raw materials met specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERAMAX CERAMIC INSERT 36 X 58
Common Device Name
PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10251901
Report Number1818910-2020-15408
Device Sequence Number689427
Product Code MRA
UDI-Device Identifier10603295012542
UDI-Public10603295012542
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K000306
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1218-87-658
Device Catalogue Number121887658
Device Lot Number9268477
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Initial Date Received by Manufacturer 06/26/2020
Supplement Date Received by Manufacturer07/13/2020
01/25/2021
Initial Report FDA Received Date07/09/2020
Supplement Report FDA Received Date07/14/2020
01/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
Patient Weight91
-
-