• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TRUMATCH CT PIN GUIDE KIT R; DRILL GUIDES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TRUMATCH CT PIN GUIDE KIT R; DRILL GUIDES Back to Search Results
Model Number 420579
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 06/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the trumatch attune ps pin guide on femoral side did not fit.Surgeon spent 5 minutes trying to fit guide on to patient femur.Guide would not fit which resulted in using manual instruments.A surgical delay of 10 minutes was noted.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned for evaluation.No evidence was found indicating product error was a contributing factor to the reported event.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUMATCH CT PIN GUIDE KIT R
Common Device Name
DRILL GUIDES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10252387
Report Number1818910-2020-15425
Device Sequence Number13574354
Product Code NJL
UDI-Device Identifier10603295430803
UDI-Public10603295430803
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420579
Device Catalogue Number420579
Device Lot Number23153
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/24/2020
Supplement Date Received by Manufacturer07/22/2020
Initial Report FDA Received Date07/09/2020
Supplement Report FDA Received Date07/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-